2-12
8753ET Specifications and Characteristics
Instrument Specifications
Table 2-9 Test Port Input
8753ET Test Port Input
Description Specification Supplemental Information
Internally Generated Harmonics (Option 002 Only)
2nd Harmonic 16 MHz to 3 GHz
at 0 dBm input level < 15 dBc, char.
at 10 dBm input level < 30 dBc, typ.
at 25 dBm input level < 45 dBc, typ.
3rd Harmonic 16 MHz to 2 GHz
at 0 dBm input level < 30 dBc, char.
at 10 dBm input level < 50 dBc, typ.
at 25 dBm input level < 50 dBc, typ.
Harmonic Measurement Accuracy
16 MHz to 3 GHz ±1.5 dB, char.
3 GHz to 6 GHz ±3 dB, char.
Harmonic Measurement Dynamic Range
Standard 40 dBc, typ.
output at 10 dBm and input
at < 15 dBm