Cary 50 Installation Instructions
9
Warm up the Cary 50
Allow half an hour for the Cary 50 instrument to warm up to ensure that the
temperature is stable. If the Cary 50 fails any of the instrument performance tests,
it may be because the temperature of the instrument has not stabilized.
Test your instrument
To test your instrument:
1 Click the Windows Start button, then All Programs > Agilent > Cary WinUV >
Validate.
2 Click Tests.
3 Click Instrument Performance Tests. Click OK.
4 Close the sample compartment. Click the Cary Start button.
5 Click Print after the tests are complete.
6 Compare your results to those supplied with the instrument.
7 Send the completed Quality card to Agilent in the supplied envelope.
Where to from here?
You will find instructions for the Cary 50 instrument and Cary WinUV software in
the Cary 50 Hardware manual or in the Cary WinUV Help.