EasyManuals Logo

Agilent Technologies E3614A User Manual

Agilent Technologies E3614A
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
A-11
STEP ACTION RESPONSE PROBABLE CAUSE
4 Check volta
g
e from pin 13 to
pin 12 of U9.
a. Measured volta
g
e is positive.
b. Measured volta
g
e is ne
g
ative.
a. Check U9A is defective.
b. Check U10 and U9D is defective.
Check R85 is open.
5 Check volta
g
e from pin 6
to pin 5 of U9.
a. Measured volta
g
e is positive.
b. Measured volta
g
e is ne
g
ative.
a. U9B is defective.
b. Check U9C is defective.
Table A-5. Low Output Volta
g
e Troubleshootin
g
(Cont’d)
Table A-6. Prere
g
ulator/Control Circuit Troubleshootin
g
STEP MEASURE RESPONSE PROBABLE CAUSE
1 Set output volta
g
e to 4.5 V +- 0.5 V for E3614A.
Set output volta
g
e to 10 V +- 1 V for E3615A.
Set output volta
g
e to 15 V +- 1 V for E3616A.
Set output volta
g
e to 26 V +- 5 V for E3617A.
2 E3614A
E3615A
Waveform form from
TP6(common) to point 6
a. Normal firin
g
pulse
b. No firin
g
pulse
a. Check CR18, CR15, Q7, Q8
for defective.
b. Proceed to step 3.
E3616A Volta
g
e from TP6
(common) to point 6
a. Hi
g
h volta
g
e (+0.7 V)
b. Low volta
g
e (0 V)
a. CR15, CR18, U2, U21
defective
b. Proceed to step 3.
3 Volta
g
e from TP6(common) to
U4 pin 1
a. Low volta
g
e (-12 V)
b. Hi
g
h volta
g
e (+5 V)
a. U3 defective
b. Proceed to step 4.
4 Volta
g
e from TP6(common) to
U5 pin 1
a. Hi
g
h volta
g
e (+15 V)
b. Low volta
g
e (-12 V)
a. U4 defective
b. Proceed to step 5.
5 Volta
g
e from pin 6 to
pin 7 of U5
a. Measured volta
g
e is positive.
b. Measured volta
g
e is ne
g
ative.
a. U5 defective
b. U6 defective
6 Set output volta
g
e to 7 V +- 1 V for E3614A.
Set output volta
g
e to 16 V +- 2 V for E3615A.
Set output volta
g
e to 25 V +- 2 V for E3616A.
Set output volta
g
e to 44 V +- 5 V for E3617A.
7 Waveform form from TP6
(common) to point 7
a. Normal firin
g
pulse
b. No firin
g
pulse
a. CR10, CR12, Q5, Q6
defective
b. Proceed to step 8.
8 Volta
g
e from TP6(common)
to U4 pin 14
a. Low volta
g
e (-12 V)
b. Hi
g
h volta
g
e (+5 V)
a. U3 defective
b. Proceed to step 9.
9 Volta
g
e from TP6(common)
to U5 pin 14
a. Hi
g
h volta
g
e (+15 V)
b. Low volta
g
e (-12 V)
a. U4 defective
b. Proceed to step 10.
10 Volta
g
e from pin 8 to
pin 9 of U5
a. Measured volta
g
e is positive.
b. Measured volta
g
e is ne
g
ative.
a. U5 defective
b. U6 defective

Questions and Answers: