Chapter 8 Specifications
Performance Specifications
150
Transient Response Time
Less than 50
m
sec for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Settling Time
Less than 90 msec for the output voltage to cha nge from 1% to 99% or vice versa
following the receipt of
VOLTage
or
APPLy
command via direct GPIB or
RS-232 interface.
OVP Accuracy, ±(% of output + offset)
<0.5% + 0.5 V
Activation time: Average time for output to start to drop after OVP condition
occurs.
<1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V