Chapter 2 General Information
Installation
33
2
Installation
Initial Inspection
When you receive your power supply, inspect it for any obvious damage that
may have occurred during shipment. If any damage is found, notify the carrier
and the nearest Agilent Technologies Sales Office immediately. Warranty
information is shown in the front of this manual.
Keep the original packing materials in case the power supply has to be returned
to Agilent Tecnologies in the future. If you return the power supply for service,
attach a tag identifying the owner and model number. Also include a brief
description of the problem.
Mechanical Check
This check confirms that there are no broken terminals or knob and that the
cabinet and panel surfaces are free of dents and scratches. Verify that the
display is not scratched or cracked.
Electrical Check
Chapter 1 describes quick operation procedure that verifies to a high level of
confidence that the power supply is operating in accordance with its
specifications. More complete verification procedures are included in the
Service Information.
Cooling and Location
Cooling
The power supply can operate at rated specifications within the temperature
range of 0 °C to 40 °C, Power supply loading is derated from 40 °C to 55 °C. A
fan cools the power supply by drawing air through the sides and exhausting it
out the back. Using an Agilent rack mount will not impede the flow of air.
Bench Operation
Your power supply must be installed in a location that allows sufficient space
at the sides and rear of the power supply for adequate air circulation. The
rubber bumpers must be removed for rack mounting.
Cleaning
No cleaning is required for this product. If you wish to remove dust from the
enclosure, use a dry cloth.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3644A and is the answer not in the manual?

Agilent Technologies E3644A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3644A
CategoryPower Supply
LanguageEnglish

Related product manuals