Programming the Status Register System
Status Groups
Chapter 3118
Standard Event Status Enable Register
The Standard Event Status Enable Register lets you choose which bits in the Standard Event
Status Register set the summary bit (bit 5 of the Status Byte Register) to 1.
*ESE <data> <data> is the sum of the decimal values of the bits you want to enable.
Example: To enable bit 7 and bit 6 so that whenever either of those bits is set to 1, the
Standard Event Status summary bit of the Status Byte Register is set to 1.
Send the command *ESE 192 (128 + 64).
Query: *ESE?
Response: Decimal value of the sum of the bits previously enabled with the
*ESE <data> command.