EasyManuals Logo

Agilent Technologies E444 AU Series Service Guide

Agilent Technologies E444 AU Series
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
Chapter 11 315
Assembly Replacement Procedures
Front Frame
Removal
1. Using a 9/16” socket drive, remove the nut securing the Ext Trigger
Input
BNC connector. Take care to not scratch the front dress panel.
2. Refer to Figure 11-5. Disconnect the ribbon cable (3) from the A2
front panel interface board. Pull the coaxial cable (4) from the front
frame and unclip from the two cable clamps.
3. Lift the front frame assembly away from the deck.
Figure 11-5 Front Frame Assembly Removal