Chapter 11 311
Assembly Replacement Procedures
Instrument Outer Case
Replacement
1. Disconnect the instrument from ac power.
2. Slide the instrument cover back onto the deck from the rear. The
seam on the cover should be on the bottom. Be sure the cover seats
into the gasket groove in the front frame.
3. Replace the four rear feet to the rear of the instrument. Torque to 21
inch pounds.
4. Using the T-20 driver, replace the handles. Torque to 21 inch pounds.
5. Replace the four bottom feet by pressing them into the holes in the
case and sliding in the opposite direction of the arrows until they
click into place. Note that the feet at the front have the tilt stands.

Table of Contents