10 Installation Note E4440-90611
Installation Procedure
Figure 5 Cable Routing
11.Reinstall the A10 3rd Converter assembly. Re-attach all cables being sure to follow
the color code on the board.
12.Locate the cable with the “60” color code in the kit. Connect the SMB end to the A10
3rd Converter J6 (60). Route the cable along the rear frame and tuck the cable under
the top brace mounting tab so the cable will lay tightly against the rear frame.
Continue routing the cable between the A31 Wideband Analog IF and A32 Wideband
Digital IF assemblies. This routing is critical. Connect the other end to the A31
Wideband Analog IF at the jack marked “60”.
13.Locate the cable with the “40” color code in the kit. Connect the SMB end to the A10
3rd Converter J4 (40). Route the cable along the rear frame. Continue routing the
cable between the A31 Wideband Analog IF and A32 Wideband Digital IF assemblies.
This routing is critical. Connect the other end of the cable to the A31 Wideband
Analog IF at the jack marked “40”.

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals