EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E444xAU Option H37

Agilent Technologies E444xAU Option H37 Manual

Agilent Technologies E444xAU Option H37
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #105 background imageLoading...
Page #105 background image
Chapter 3 105
Making Measurements
Modulation Distortion Measurement
Making Measurements
3. Press FM De-Emphasis to select a value. This key is enabled only when the
Modulation Mode is FM.
4. Press
Modulation Mode to select from AM, FM and PM.
Step 9. To adjust the display, press
AMPLITUE/Y Scale, then:
1. Press the
Display Unit key and select a display unit key. The default
setting is
%.
2. Press the
Display Mode key to toggle between Normal and Ratio. The default
setting is
Normal.
3. Press the
Ratio Ref key to toggle between Auto and Man. When you select Man,
input the reference value and unit. When the reference is set to 0 or the
measurement result is 0 under log ratio mode, the result displayed is ā€œ- - -ā€.
4. Press the
Ratio Mode key to toggle between Log and Linear.
Modulation Distortion Measurement Example
Figure 3-17 AM Modulation Distortion Measurement Result
Signal Under Test: RF Frequency: 1.25 GHz; Amplitude: 0.0 dBm;
AM Depth: 10%; AM Rate: 100 kHz; Modulation: On.

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals