Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
145
Front-Panel Key and SCPI Command
Reference
4.2.5.3 Int Preamp
Sets the state of the Internal Preamplifier.
4.2.5.4 Microwave Preselector
Sets the state of the Microwave Preselector.
Mode: MRECEIVE
Key Path:
Input, Internal Preamplifier
Remote Command: [:SENSe]:POWer:RF:GAIN[:STATe] ON|OFF|0|1
[:SENSe]:POWer:RF:GAIN[:STATe]?
Preset: OFF
State Saved: Saved in instrument state.
Range: On | Off
Remote Command Notes: Global to the current mode.
Example: POW:RF:GAIN ON
POW:RF:GAIN?
Mode: MRECEIVE
Key Path:
Input, Microwave Preselector
Remote Command: [:SENSe]:POWer:RF:MW:PRESelector[:STATe] ON|OFF|0|1
[:SENSe]:POWer:RF:MW:PRESelector[:STATe]?
Preset: ON
State Saved: Saved in instrument state.
Range: On | Off
Notes: This key is grayed out when center frequency <= 3.05 GHz, and the state will
always be ON.
Remote Command Notes: Global to the current mode.
Example: POW:RF:MW:PRES ON
POW:RF:MW:PRES?

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals