EasyManuals Logo

Agilent Technologies E444xAU Option H37 Manual

Agilent Technologies E444xAU Option H37
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #209 background imageLoading...
Page #209 background image
Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
209
Front-Panel Key and SCPI Command
Reference
4.4.5.1 SPAN/X Scale
The SPAN/X Scale key accesses the menu to set the desired horizontal scale and associated settings of the
waveform.You must have the Demod Waveform selected under the Trace/View key.
4.4.5.1.1 Scale/Div
This key is for Scale/Div control
Table 4-18 FM Deviation Numeric Results Window
Name Corresponding Results
FM Deviation N=1 1st
FM Deviation output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Mode: MRECEIVE
Key Path:
Span X Scale
Remote Command: :DISPlay:FMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision <seconds>
:DISPlay:FMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision?
Unit: Seconds
Preset: 100us
State Saved: Saved in instrument state.
Min: –10 s
Max: 1 s
Dependencies/Couplings: See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:FMD:WIND:TRAC:X:PDIV 100us
:DISP:FMD:WIND:TRAC:X:PDIV?

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals