4.4.5.4.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
Key Path: Meas Setup
Remote Command: [:SENSe]:FMDeviation:TRIGger:SOURce
IMMediate|IF|EXTernal[1]|EXTernal[2]|RFBurst
[:SENSe]:FMDeviation:TRIGger:SOURce?
Default: IMMediate
State Saved: Saved in instrument state.
Range: Free Run (Immediate)|Video (IF Envlp)|RF Burst|Ext Front|Ext Rear
Example: :FMD:TRIG:SOUR IF
:FMD:TRIG:SOUR?
Key Path: Meas Setup
Remote Command: [:SENSe]:FMDeviation:CAPTure[:TIME] <time>
[:SENSe]:FMDeviation:CAPTure?
[:SENSe]:FMDeviation:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:FMDeviation:CAPTure:AUTO?
Preset/Default: 250 ms
State Saved: Saved in instrument state.
Min: 1 μs
Max: 10 s
Resolution 1 ms
Example: :FMD:CAPT 0.25
:FMD:CAPT?
:FMD:CAPT:AUTO OFF
:FMD:CAPT:AUTO?

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals