Chapter 3 25
Operation Verification
Checking the Output Power
Problems with Output Power
If a problem occurs when verifying output power levels, perform the following procedure:
1. Verify the power sensor being used in this test is the appropriate sensor.
2. Verify the power sensor calibration factors, if required, are correct.
3. Verify the power sensor is properly calibrated to the power meter.
If the problem continues, perform a Power Level Accuracy performance test. Refer to the Calibration
Guide.