Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 103
Programming Examples
LAN Programming Interface Examples
// Open SICL instrument handle using VXI-11 protocol
sprintf(instNameBuf, "lan[%s]:inst0", instrumentName);
id = iopen(instNameBuf);// Open instrument session
itimeout(id, 1000);// Set 1 second timeout for operations
printf("Setting frequency to 1 Ghz...\n");
iprintf(id, "freq 1 GHz\n");// Set frequency to 1 GHz
printf("Waiting for source to settle...\n");
iprintf(id, "*opc?\n");// Query for operation complete
iscanf(id, "%d", &opcResponse); // Operation complete flag
if (opcResponse != 1)// If operation fails, prompt user
{
printf("Bad response to 'OPC?'\n");
iclose(id);
exit(1);
}
iprintf(id, "FREQ?\n");// Query the frequency
iscanf(id, "%t", &buf);// Read the signal generator frequency
printf("\n");// Print the frequency to the display
printf("Frequency of signal generator is %s\n", buf);
ipromptf(id, "*IDN?\n", "%t", buf);// Query for id string
printf("Instrument ID: %s\n", buf);// Print id string to display
iclose(id);// Close the session
return 0;
}
VXI-11 Programming Using VISA and C++
The following program uses the VXI- 11 protocol and the VISA library to control the signal generator.
The signal generator is set to a –5 dBm power level and queried for its ID string. Before running this
code, you must set up the interface using the Agilent IO Libraries IO Config utility. vxivisa.cpp
performs the following functions:
• sets signal generator to a –5 dBm power level
• queries signal generator for an ID string
•error checking
The following program example is available on the signal generator Documentation CD- ROM as
vxivisa.cpp.
//****************************************************************************************
// PROGRAM FILE NAME:vxivisa.cpp
// Sample test program using the VISA libraries and the VXI-11 protocol
//

Table of Contents

Other manuals for Agilent Technologies E8257D

Questions and Answers:

Agilent Technologies E8257D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8257D
CategoryInverter
LanguageEnglish

Summary

Getting Started with Remote Operation

Interfaces

Details various interface types (GPIB, LAN, RS-232, USB) for instrument communication.

IO Libraries and Programming Languages

Covers IO libraries and common programming languages for instrument control.

Using IO Interfaces

Using GPIB

Details GPIB interface for instrument connection and control, including setup and verification.

Using LAN

Explains LAN interface protocols (VXI-11, Sockets, Telnet, FTP) for remote instrument programming.

Programming Examples

Using the Programming Interface Examples

Introduces programming examples for remote control using various interfaces and languages.

GPIB Programming Interface Examples

Provides examples for GPIB programming and instrument control, including interface checks and queries.

LAN Programming Interface Examples

Covers LAN programming examples using VXI-11 and Sockets LAN protocols for instrument control.

Programming the Status Register System

Overview

Provides an overview of the status register system for monitoring signal generator conditions and errors.

Status Register Bit Values

Explains how each bit in a register is represented and how to enable/query bits using SCPI commands.

Accessing Status Register Information

Details methods for monitoring conditions, including polling and service request interrupts.

Status Byte Group

Describes the Status Byte Group, including the Status Byte Register and Service Request Enable Register.

Creating and Downloading Waveform Files

Creating and Downloading User-Data Files