Service Guide E8364-90038 3-35
PNA Series Microwave Network Analyzers      Tests and Adjustments
E8362C, E8363C, E836C Performance Tests (Agilent N7840A Software Package)
Calibration Coefficient Test
Function of the Test: To verify the uncorrected calibration coefficients of your analyzer. The calibration 
coefficients are measured in forward and reverse direction. Refer to “Error Terms” on page 8-3 for error term 
information relating to the calibration coefficients measured.
Specification Tested: Uncorrected System Performance
Equipment Used: A calibration kit and a test cable.
Description of the Test:
1. A series of 2-port calibrations are performed. Two full SOLT 2-port calibrations are performed on each 
port. Isolation is turned off during each 2-port calibration.
2. A test cable is attached to the first port, and a calibration is performed at the end of the cable and at the 
second port. The cable is moved to the second port and another calibration is performed. Using two 
calibrations helps to eliminate the unknown characteristics of the through cable.
3. Each calibration produces 12 error terms.
4. The error terms are used to determine the following characteristics: directivity, source match, load 
match, reflection tracking, and transmission tracking.
If the Analyzer Fails this Test:
• If the analyzer fails tracking error terms only, perform “Receiver Calibration Adjustment” on page 3-47 
and repeat this test.
• Failure of any other error terms indicate a hardware failure. Refer to the appropriate error term 
discussion in Chapter 8, “General Purpose Maintenance Procedures” for a typical cause of failure. Refer 
to Chapter 7, “Repair and Replacement Procedures,” for instructions on replacing the suspected faulty 
component or assembly.