1-4 Chapter 1
Installing Your Analyzer
STEP 2. Meet Electrical and Environmental Requirements
STEP 2. Meet Electrical and Environmental Requirements
1. Ensure the available AC power source
meets the following requirements:
2. Ensure the operating environment meets
the following requirements:
CAUTION This instrument has
autoranging line voltage
input. Be sure the supply
voltage is within the specified
range.
The specified range for the power
requirements is:
•90 132 V
AC
•48 66 Hz / 400 Hz (single phase)
- or -
198 264 V
AC
•48 66 Hz (single phase)
The instrument power consumption is
600 VA maximum.
•0 to 40 °C
5% to 95% relative humidity at 40 °C
(non-condensing)
0 to 4,500 meters (0 to 14,760 feet)
altitude
NOTE Some analyzer performance
parameters are specified for
25 °C. Later, you can refer to
the Help system in the
analyzer for the complete
specifications.
3. Verify that the AC power cable is not damaged, and that the power-source outlet
provides a protective earth contact.
CAUTION Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord may cause
product damage.

Other manuals for Agilent Technologies E8362C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8362C and is the answer not in the manual?

Agilent Technologies E8362C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8362C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals