24 Installation Note E8364-90024
Step 13. Perform Post-Upgrade Adjustments and Calibration
The following adjustments must be made due to the change in the full frequency range of the
analyzer.
source calibration
receiver calibration
These adjustments are described in the PNA service guide and in the PNA on-line HELP. A
list of equipment required to perform these adjustments can be found at “Equipment
Required for Post-Upgrade Adjustments” on page 6.
Performance Tests and System Verification
The analyzer should now operate and phase lock over its entire frequency range.
If you experience difficulty with the basic functioning of the analyzer, contact Agilent. Refer to
“Contacting Agilent” on page 3.
Although the analyzer functions, its performance relative to its specifications has not been
verified for the additional frequency range enabled by this upgrade.
It is recommended that a full instrument calibration be performed using the PNA
performance test software.
If the testing of the analyzers full range of specifications is not required, a system verification
can be performed.
Refer to the analyzers service guide for information on performance tests and system
verification.

Other manuals for Agilent Technologies E8364B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8364B and is the answer not in the manual?

Agilent Technologies E8364B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8364B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals