Installation Note E8364-90024 7
Step 1. Remove the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front handles
in the vertical position. Do not place the analyzer on its front panel without the
handles. This will damage the front panel assemblies.
Refer to Figure 2 for this procedure.
1. Disconnect the power cord (if it has not already been disconnected).
2. With a T-20 TORX driver, remove the four rear panel feet (item
) by removing the center
screws (item
).
3. Slide the four bottom feet (item
) off the cover.
4. With a T-20 TORX driver, remove the strap handles (item
) by loosening the screw
(item
) on each end until the handle is free of the analyzer.
5. Slide the cover off of the frame.
Figure 2 Outer Cover Removal
NOTE
The figure above shows the E8363/4B front panel and floppy disk drive.
The E8363/4C front panel has a slightly different appearance and does not
include a floppy disk drive.