3-14 Chapter 3
Making GSM Base Station Measurements
Making the Power Versus Time Measurement
Making the Measurement
1. Press the Measure key.
2. Press the
Power vs Time key.
Factory default parameter settings provide a GSM compliant
measurement. For special requirements, you may need to change
default settings:
• Mode setup and frequency/channel parameters.
Use the
Mode Setup and Frequency Channel keys to change these
parameters for all measurements made within the current mode. For
further information refer to Chapter 1 of this document.
• Measurement setup parameters.
These are measurement specific parameters changed using the
Measurement setup (
Meas Setup) menu. Parameters can be returned
to default settings at any time by pressing
Meas Setup, More (1 of 2)
and
Restore Meas Defaults. For further information on measurement
setup parameters, refer to Chapter 5 of the ESA-E Series Spectrum
Analyzers GSM Measurement Personality User’s Guide.
NOTE Parameters that exist under the Meas Setup Advanced key seldom need
to be changed. Any changes from the default values may result in
invalid measurement data.
Power Time Custom Masks
For the Power vs Time measurement, you can define a user
configurable limit mask to apply to the measured burst. This feature
can only be accessed via SCPI commands. For further information refer
to the ESA-E Series Spectrum Analyzers GSM Measurement Personality
Programing Guide.