8 Agilent 1200 Series Valves
1 Installation and Configuration
Site Requirements
Site Requirements
The requirements for the Agilent 1200 Series Valves are displayed Table 1 and
Table 2 on page 8. The 24 Volts DC power can be supplied by an external
power supply (0950-4422) or by one of the following Agilent 1200 Series LC
modules with DC CAN adapter.
Agilent 1200 Series preparative pump (G1361A)
Agilent 1200 Series fraction collectors (G1364A, G1364B, G1364C, G1364D)
Agilent 1200 Series high performance autosampler(G1367B)
Agilent 1200 Series micro well-plate autosampler (G1377A)
Agilent 1200 Series dual-loop autosampler PS (G2258A)
Table 1 Site Requirements for Agilent 1200 Series Valves
Type Specification
Weight 1.9 Kg (4.2 lbs)
Dimensions
(height x width x depth)
92 x 84 x 200 mm
(9.2 x 3.3 x 8.0 inches)
Power supply 24 Volts DC (1.7 amps)
Ambient operating temperature 4 to 55°C (39 to 131°F)
Humidity < 95%, non-condensing
Safety Standards IEC, CSA, UL, EN
Installation category II,
pollution degree 2
For indoor use only
Table 2 System Requirements
Revision
Agilent 1200 Firmware A.05.04 and higher
Control Module Firmware G1323B B.03.11 and higher
Agilent ChemStation Software A.09.03 and higher