Error Messages
Error Message Listing
130
The microprocessor immediately switches off the heater associated with the
failed sensor. You must correct the problem before you can continue to use the
sampler.
CAROUSEL RESET
Probable Causes
The sampler was reset.
Suggested Actions
See “Agilent 7694 or 7694 RESET Agilent 7694E or 7694E RESET” on
page 128.
PRESS STOP TO CONFIRM
The sampler is waiting for confirmation before changing a setpoint.
RAM CHECK ERR
PRESS CLEAR TO RESET
Probable Causes
The RAM check value is different from the stored value.
Suggested Actions
Press Clear to continue. The instrument is stopped and you must totally reset
it. This resets the microprocessor and automatically initializes all the data
(method parameters, advanced function values, flags and all the variables)
with default values. If the unit fails to reset, try turning the power off, then on.
If this fails, contact Agilent service.