Error Messages
Error Message Listing
132
TRAY RESET
Probable Causes
The sampler was reset.
Suggested Actions
See “Agilent 7694 or 7694 RESET Agilent 7694E or 7694E RESET” on
page 128.
UP MOTOR ERROR
Probable Causes
The elevator assembly that lifts the vial onto the sampling needle is unable to
reach the needle.
Suggested Actions
Remove all vials from the oven and the gripper. Power cycle the headspace
sampler and hold down the clear key until the tray is reset.
VIAL NOT FOUND
Probable Causes
The robotic arm is unable to retrieve a vial from the oven
The headspace sampler is unable to load a vial into the oven
Suggested Actions
1. Manually remove any vials that are in the oven carousel.
WARNING Vials in the oven may be hot.