7 General Maintenance
Maintaining the Analyzer
174 5977B Series MSD Operating Manual
Maintaining the Analyzer
Scheduling
None of the analyzer components require periodic maintenance. Some tasks,
however, must be performed when MSD behavior indicates they are necessary.
These tasks include:
Cleaning the ion source
Replacing filaments
Replacing the electron multiplier horn
The Agilent 5977B Series MSD Troubleshooting and Maintenance Manual
provides information about symptoms that indicate the need for analyzer
maintenance. The troubleshooting material in the online help in the Agilent
MassHunter GC/MS Acquisition software provides more extensive information.
Precautions
Cleanliness
Keep components clean during analyzer maintenance. Analyzer maintenance
involves opening the analyzer chamber and removing parts from the analyzer.
During analyzer maintenance procedures, take care to avoid contaminating the
analyzer or the interior of the analyzer chamber. Wear clean gloves during all
analyzer maintenance procedures. After cleaning, parts must be thoroughly
baked out before they are reinstalled. During and after cleaning, analyzer parts
should be placed only on clean, lint-free cloths.
CAUTION
If not done correctly, analyzer maintenance can introduce contaminants into
the MSD.
WARNING
The analyzer operates at high temperatures. Do not touch any part until you
are sure it is cool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7079B and is the answer not in the manual?

Agilent Technologies G7079B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7079B
CategoryWater Pump
LanguageEnglish

Related product manuals