EasyManuals Logo

Agilent Technologies G7081B User Manual

Agilent Technologies G7081B
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
6 Operating in CI Mode
To Verify NCI Performance
164 5977B Series MSD Operating Manual
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs only.
Materials needed
OFN, 100 fg/µL 3 × 1 mL (5188-5347)
Procedure
1 Verify that the MSD performs correctly in EI mode.
2 Load the NCICH4.U tune file, and accept the temperature setpoints.
3 Select Gas A and set flow to 40%.
4 In Tune and Vacuum Control view, run CI Autotune. (See “To Perform an NCI
Autotune (Methane Reagent Gas)” on page 161.)
Note that there are no criteria for a “passing” Autotune in CI. If the Autotune
completes, it passes.
5 Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6 Verify that the system conforms to the published sensitivity specification.
Please refer to the Agilent Web site at www.agilent.com/chem for
specifications.
CAUTION
Always verify MSD performance in EI before switching to CI operation. (See
“To Verify EI System Performance” on page 131.) Always set up the CI MSD in
PCI first, even if you are going to run NCI.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7081B and is the answer not in the manual?

Agilent Technologies G7081B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7081B
CategoryWater Pump
LanguageEnglish

Related product manuals