110 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
AIV Fuse
Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the
inlet-valve electronic fuse to open.
Probable cause Suggested actions
1
Defective active inlet valve.
Restart the module. If the error message
appears again, exchange the active inlet valve.
2
Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.