54 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
3
Using the Pump
Best Practices
Best Practices
Daily / Weekly tasks
Daily tasks
Replace mobile phase based on water/buffer.
Replace organic mobile phase latest every second day.
Check seal wash solvent.
Weekly tasks
Change seal wash solvent (10 % / 90 % isopropanol/water) and bottle.
Flush all channels with water at 2.5 – 3 mL/min for 5 min to remove salt
deposits if buffer applications were used.
Inspect solvent filters for dirt or blockages. Clean or exchange if no flow is
coming out of the solvent line when removed from the degasser inlet.
Power up / Shut-down the pump
Power up the pump
Use new or different mobile phase (as required).
Purge pump heads with 2.5 – 3 mL/min for 5 min.
Stabilize the system by running for 10 – 20 min.
Long-term shut-down of the system
Flush system with water to remove buffer.
Remove all samples from the sampler and store according to good
laboratory practice.
Use recommended solvents to store the system.
Power off the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7110B and is the answer not in the manual?

Agilent Technologies G7110B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7110B
CategoryWater Pump
LanguageEnglish

Related product manuals