EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7110B

Agilent Technologies G7110B User Manual

Agilent Technologies G7110B
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 245
LAN Configuration
12
Manual Configuration
5 Use the “/” and press Enter to list the current settings.
6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.
Figure 33 Telnet - Current settings in "Using Stored" mode
information about the LAN interface
MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings
TCP/IP status - here ready
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected
Figure 34 Telnet - Change IP settings
change of IP setting to
Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7110B and is the answer not in the manual?

Agilent Technologies G7110B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7110B
CategoryWater Pump
LanguageEnglish

Related product manuals