68 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
3
Using the Pump
Setting up the Pump with the Instrument Control Interface
Pressure Limits Max: 600 bar
(8700 psi) for flow
rates up to
5mL/min. 200bar
(2900 psi) for flow
rates > 5 mL/min.
Min: any value
between 0and the
upper pressure limit
setting.
Sets the maximum and minimum pressure limits for the pump.
•Max is the maximum pressure limit at which the pump will switch
itself off, protecting the analytical system against over-pressure.
•Min is the minimum limit at which the pump will switch itself off, for
example, if any solvent reservoir is empty, this prevents system
damage by pumping air.
Timetable See “Timetable Settings” on page 70
Advanced See “Advanced Settings” on page 69
Table 1 2 Method parameters
Parameter Limits Description