Agilent Technologies HP 83711A
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
L
VL
CR
The
LVL
CR (level
correct)
function
allows
you
to
calibrate
external
path
losses
in the
signal path.
The
level
correct
function
creates
a
table
in
synthesizer
memory of
external path
loss values
versus
frequency
.
Invoking
the L
VL CR
(level correct)
function
(pressing
the
4
SHIFT
5
key and
then
the
4
INT
LEVEL
5
key) causes
the synthesizer
to prompt
you
for
the
start
frequency
, stop
frequency
,
and
number
of
points
in
the
level
correction
table
.
Once
the start
frequency
,
stop
frequency
,
and
number
of
points
are
entered,
you
are
given
the
option
of
running
the
level
correct
routine
.
When
4
SHIFT
5
,
4
INT
LEVEL
5
(
L
VL
CR
)
is
pressed,
the
synthesizer
initially
prompts
you
for
the
level
correct
start
frequency
.
When
4
SHIFT
5
,
4
INT
LEVEL
5
(
LVL
CR
)
is pressed,
the text
COR
START
XXXGHZ
will
be displayed
where
XXX
is
the
current
start
frequency
.
At
this
point,
a
new
start
frequency can
be
entered
or
the
existing
start
frequency
can
be
modied
using
the
knob
or
arrow
keys
.
The
start
frequency
is
used
to
determine
the
beginning
frequency
for
the
level
correction table
.
The
start
frequency
is
the
rst
frequency
point
in
the
table.
N
OT
E
Y
ou
do
not
have
to
choose
a
new
start
frequenc
y
(or
stop frequenc
yor
number of
points).
T
o
keep
the
same
parameter
value
,
press
4
Hz
5
(ENTER)
without
entering
a
new
parameter
and
the
next
parameter
in
the
sequence
will
be
displa
yed.
6-31