Error
Messages
Messages
See
error
1803.
0
311
0
311,"Memory
error;(
0
311)"
See
error
1803.
0
310
0
310,"System
error;(
0
310)"
Some
problem
occurred
while
parsing
an
HP-IB
command
or
query
.
Insure
that
your
programming
is
correct
and
try
the
command
again.
0
300
0
300,"Device
specic
error;(
0
300)"
A
remote
command
or
query
could
not
be
executed
because
an
error
occurred
in
the
synthesizer
.
0
278
0
278,"Macro
header
not
found;(
0
278)"
A
*GMC?
or
*RMC
macro
label
could
not
be
found
in
the
list
of
dened
macro
labels
.
Use
*LMC?
to
get
a
list
of
all
the
currently
dened
macro
labels
.
0
277
0
277,"Macro
redenition
not
allowed;(
0
277)"
Indicates
that
a
macro
label
in
the
*DMC
command
could
no
be dened
because
the
macro
label
was
already
dened.
0
276
0
276,"Macro
recursion error;(
0
276)"
The
nesting/recursion
of
macros
is
deeper than
4
levels
.
Don't
use
more
than
4
levels
when
dening macros
of macros
.
0
275
0
275,"Macro
denition
too long;(
0
275)"
The
macro denition
must
be
255
characters
or
less
.
0
274
0
274,"Macro
parameter
error;(
0
274)"
A
macro parameter
placeholder was
improperly
used.
0
273
0
273,"Illegal macro
label;(
0
273)"
Indicates
that
a
macro
label
dened
in
the
*DMC
command
has
a
legal
string syntax; but, it is too long. It is the
same as a common command
header, or contain invalid header syntax.
0
272
0
272,"Macro execution error;(
0
272)"
Indicates that a syntactically legal macro program data sequence could
not be executed due to some error in the macro denition.