EasyManua.ls Logo

Agilent Technologies HP 83711A - Page 217

Agilent Technologies HP 83711A
251 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
Messages
Messages
The
above
command
is
recognized
by
the
synthesizer
in
HP
8673
emulation
mode
,
but
it
is
not
emulated.
The
command
is
thus
discarded
and
no
action
is
taken.
Refer
to
Chapter
4
of
the
HP
8673
Compatibility
Guide
for
more
information.
1540
0
301,"8673
command
SP
not
emulated
(1540)"
The
above
command
is
recognized
by
the
synthesizer
in
HP
8673
emulation
mode
,
but
it
is
not
emulated.
The
command
is
thus
discarded
and
no
action
is
taken.
Refer
to
Chapter
4
of
the
HP
8673
Compatibility
Guide
for
more
information.
1541
0
301,"8673
command
IF
not
emulated
(1541)"
The
above
command
is
recognized
by
the
synthesizer
in
HP
8673
emulation
mode
,
but
it
is
not
emulated.
The
command
is
thus
discarded
and
no
action
is
taken.
Refer
to
Chapter
4
of
the
HP
8673
Compatibility
Guide
for
more
information.
1802
1802,"Low
Battery
V
oltage;(1802)"
The
synthesizer
battery
voltage
is
low
.
This
could
cause loss
of RAM
data
if
signal
generator
power
is
turned
o.
Note:
Calibration
data
will
never
be
lost.
1803
1803,"RAM
data
lost
at power
on;(1803)"
All
RAM
data
was
lost.
This includes
all
front
panel
settings
,
save/recall
registers
,
level
corrections
, and
other user
settable
values
.
This
error
message
can
occur
when the
battery voltage
is
low
,
or
options
change
in
the
synthesizer
. Note:
Calibration data
will
never
be
lost.
1804
1804,"Self test
failure
,
run
the
self
test;(1804)"
The
power-on self
test detected
an
error
or
warning.
See
the
explanation
for error
number 4000.
1805
1805,"Processor
Board
or
IBUS
test
F
ailure;(1805)"
The power-on self test detected an error or warning for
the
microprocessor board circuits or power supply monitors
. See the
explanation for error number 4000.
1806
1806,"ROM checksum test failure;(1806)"
The synthesizer ROM check sum does not match the data in ROM. See
the explanation for error number 4000.
8-29