EasyManua.ls Logo

Agilent Technologies HP 83711A - Verifying Synthesizer Functionality

Agilent Technologies HP 83711A
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V
erifying
Synthesizer
Functionality
The
verication procedure
is suitable
for
incoming
inspection;
however
,
you
can refer
to the
HP
83711A/12A
and
83711B/12B
Synthesized
CW
Generators
Service Guide
or the
HP 83711A/12A,
HP
83711B/12B
,
HP
83731A/32A,
and HP
83731B/32B Synthesized
CW Generators
Calibration
Guide
for procedures
that test
all warranted
specications
.
1.
Activate
the
SELF
TEST
special
function.
T
o
activate
the
SELF
TEST
special
function,
perform
the
following
procedure
.
a.
Press
the
4
SPCL
5
key
.
b.
Press
4
5
5
on
the synthesizer
numeric keypad.
c.
T
erminate
the
special function
entry by
pressing
the
4
Hz
5
(ENTER)
key
.
The
display
will
read
SELF TEST?,
PRESS ENTER
.
2.
Press
the
ENTER
key
again
to
initiate
the
synthesizer
self
test
routine
.
When
the
self
test
routine
is
running,
the
display
will
alternately
read
SELF TESTING
and
PRESETTING
INSTRUMENT
.
After
the
test
completes
,
the display
momentarily
reads
SELF
TEST
PASSED
.
3.
If
the self
test
indicates
an
error
condition,
refer
to
the
section
entitled
\
If
Y
ou
Encounter
a
Problem
"
at
the
end
of
this
chapter
.
If
an
error
condition
exists
,
the
display
will
momentarily
read
TEST
XX
=
YY
where
XX
and
YY
are
numbers
indicating
the
error
condition.
1-7