1Getting Started
18 N1913/1914A Installation Guide
5 Connect the sensor to the POWER REF connector.
6 When the sensor is connected to the POWER REF, you can zero and calibrate the
measurement path quickly by pressing , . On dual channel
meters, press or .
The Zeroing pop-up is displayed during the zeroing process, the Calibrating pop-up during
calibration. The meter-sensor measurement path is calibrated when the Calibrating pop-up
disappears.
Zero + Cal
Zero + Cal A
Zero + Cal B