Using Your Column
Installation
• The direction of ow is marked on the column.
• 1.8 µm columns (ZORBAX RRHT, ZORBAX RRHD) can only
be operated in the direction ow marked on the column.
For removable, zero-dead volume column connections
Agilent recommends the use of the InnityLab Quick Connect
and Quick Turn family. Choices are:
Maximum
System
Pressure Recommended Fitting Part Numbers
Up to 400 bar InnityLab Quick Turn Fitting
(nger-tight)
Fitting: 5067-5966
Up to 800 bar InnityLab Quick Turn Fitting
(with Mounting tool)
Fitting: 5067-5966
Mounting Tool:
5043-0915
Up to 1300 bar InnityLab Quick Connect Fitting Fitting: 5067-5965
For more information and part numbers, please see to the
AgilentInnityLab Fitting Brochure (5991-5164EN).
ENGLISH
InnityLab Quick Connect
assembly, p/n 5067-5961
InnityLab Quick Turn
tting, p/n5067-5966
Learn more at: www.agilent.com/chem/innitylabttings