EMC Information
Immunity
This control unit has been stringently tested for susceptibility to electromagnetic radiation over the
frequency range 80 MHz to 2.5 GHz. The test was conducted on this control unit and passed the
requirements of IEC 60601-1-2:2007.
The control unit is intended for use in the electromagnetic environment specified below. The customer or
user of the control unit should ensure that it is used in such an environment.
Table 202 Guidance and Manufacturer's Declaration - Immunity All Equipment and Systems
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment Guidance
ESD
IEC 61000-4-
2
±6kV Contact
±8kV Air
A
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the r/h
should be at least 30%
EFT
IEC 61000-4-
4
±2kV Mains
±1kV I/Os
A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-
5
±1kV Differential
±2kV Common
A
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage
Dips/Dropout
IEC 61000-4-
11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
A
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the control unit
requires continued operation during power
mains interruptions, it is recommended that
the control unit be powered from an
uninterruptible power supply or battery.
Power
Frequency
50/60Hz
Magnetic
Field
IEC 61000-4-
8
3A/m A
Power frequency magnetic fields should be
that of a typical commercial or hospital
environment.