EMC Information
Equipment and Systems which are NOT Life-supporting
Recommended Separation Distances for the control unit
The control unit is intended for use in the electromagnetic environment in which radiated disturbances
are controlled. The customer or user of the control unit can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF Communications Equipment and the
control unit as recommended below, according to the maximum output power of the communications
equipment.
D = (3.5/V1) (S
QRT P)
D = (3.5/E1) (SQRT P)
80 to 800 MHz
D = (7/E1) (SQRT P)
800 MHz to 2.5 GHz
Table 206 Recommended Separation Distances between portable and mobile RF
Communications equipment and the control unit.
Compliance Level
Cond RF
3
Rad RF-800MHz
3
Rad RF - 2.5GHz
3
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33