Device Operation Specications
Overview
Primary Operating Mode
The AIROS 8 can be used in two primary Operating Modes: Gradient Mode or
Pressure Mode.
Gradient Mode: User can set the starting pressure in Chamber 1 on the
garment, the most distal chamber. Pressure decreases at a xed gradient of 4%
moving distally to proximally.
Pressure Mode: User can set the pressure in each of the 8 chambers in the
garment. The default gradient is 4%. Once a setting is changed in a chamber,
the following chamber pressures will automatically be updated to a value
equaling a 4% pressure gradient.
NOTE: User may operate the AIROS 8 using the factory default
settings. The factory default settings are Gradient Mode operation
with 45mmHg of pressure in the distal chamber; a 60-minute therapy
time; a 60-second cycle time; Lymphedema Preparation Treatment
(LPT) Mode set to OFF; and no chambers skipped.
PRIMARY OPERATING MODE SETTINGS GRADIENT MODE PRESSURE MODE
Pressure Range (set in distal chamber) 30-80mmHg (increments of 1mmHg) 35-80mmHg (increments of 1mmHg)
Cycle Time 40 or 60 seconds 40 or 60 seconds
Deation 12 seconds 12 seconds
Therapy Time 15, 30, 45 or 60 minutes 15, 30, 45 or 60 minutes
Gradient 4% (xed) 4% (default; can be changed by user)
Skip Chamber Option Yes (1 chamber) Yes (1 chamber)
Secondary Lymphedema Preparation Treatment
(LPT) Mode Option
Focal or Clearing Focal or Clearing
SECONDARY LYMPHEDEMA PREPARATION
TREATMENT (LPT) MODE SETTINGS
FOCAL MODE CLEARING MODE
Pressure Options 40, 45 or 50mmHg 15 or 20mmHg
Chamber Selection 1-4, 2-5, 3-6, 4-7 or 5-8 -
Therapy Time 10 or 15 minutes -