Attaching and Detaching Garment
Connector
The Garment Connector on the garment connects easily to the Garment Connector
Port on the Front Panel of the device.
While holding the Garment Connector in front of the device, the hose that is
colored black should be on the left hand side. Plug the prongs on the Garment
Connector into the holes on the Garment Connector Port on the device.
The Garment Connector can also be easily detached from the device. Simply pull
out the Garment Connector with a light force and the garment will detach from the
device.
Home users are not recommended to plug the garments into the device. It is
recommended to have a medical professional plug the garments into the device.
Fuse Replacement
The safety fuse on the back panel of the device can sometimes blow for different
reasons such as a power surge or the normal aging of the electronic components.
The safety fuse is located in between the POWER PLUG and the MAIN POWER
SWITCH.
When occasional fuse damage does happen, a medical professional can replace
the fuse as long as a part that has the following parameters is ordered (2.5 AMP
T250VACH SLO-BLO).
Prior to removal of fuse, disconnect the power cord. While pushing inward on
fuse cap, turn counterclockwise to release cap and remove fuse. After placing the
new fuse in the cap slot, push cap and fuse inward and turn clockwise to secure.
NOTE: The outer safety fuse is the only item serviceable by someone
other than an AIROS Medical trained technician. AIROS Medical
trained technicians have been trained specically for the manufacture
and repair of all AIROS Medical devices.