◇ X-ray tube ; Fixed anode tube
- Model : D-125
- Focal spot : 1.2 mm
- Target angle : 16 degrees
- Anode heat storage capacity : 50 kHU
- Weight : Approx. 350g
- Filtration : min. 0.8 mmAleq.
◇ Beam limiting device
- Model : BLD-150AJ
- Lamp : DC12V / 100W Halogen
- Filtration : min. 1.3 mmAleq.
- Min. X-ray field size : ≤ 5cm x 5cm at SID 100cm
- Max. X-ray field size : ≤ 35cm x 35cm at SID 65cm
- Manually adjustable, full field light beam collimator.
- Weight : 3.3 kg
◇ Weight : 14.5 kg
◇ Environmental conditions ;
- Operation : Temperature 5 ∼ 40 ℃
Relative humidity 30 ∼ 85 %
Atmospheric pressure 600 ∼ 1060 hPa
- Storage : Temperature -40 ∼ 70 ℃
Relative humidity 10 ∼ 95 %
Atmospheric pressure 50 ∼ 1060 hPa
-------------------------------------------------------------
AJEX240H / Operating instructions -34-
AXR-240H-MN01 (E) / Rev. 03