7. ERROR CODE
There are five types of errors ;
The following error codes are shown in the mAs display on the display
panel.
Error code
- E01 : Main switching overload error
- E02 : kV or mA overload error
- E03 : kV or mA feedback undetected error
- E04 : Unintended X-ray exposure finish error
- E05 : Overheat error ( single tank, Heat sink )
When error codes are displayed, "Error" message is flickered on the kV
display and flicker buzzer sound is occurred.
If error code is displayed, perform the following checks.
1) ERROR code E01 ; Main switching overload error
a) Press enter key, then error mode will be returned normal radiographic
mode. If error mode is not returned normal radiographic mode,
contact AJEX immediately.
b) Releasing an exposure.
If error code E01 is displayed again on the mAs display, contact
AJEX immediately.
Refer to page 28. 6-3. Service
-------------------------------------------------------------------------
AJEX240H / Operating instructions -30-
AXR-240H-MN01 (E) / Rev. 03