8. Technical data
◇ Line input power : 110V or 220V∼, 50/60Hz, Single phase
◇ Input power : 5.1 kVA max.
◇ Maximum output power: 2.4 kW
◇ Generator type : High frequency inverter type, main switching frequency
- Approx. 50kHz
◇ Rating ;
- kV range : 40 ∼ 120 kV ; 1kV step increment
- mAs range : 0.4 ~ 120 mAs
- Exposure time : 0.02 ∼ 4.0 sec.
0.8 ∼ 120 mAs at 40 ~ 60 kV
0.6 ∼ 100 mAs at 61 ∼ 70 kV
0.6 ∼ 80 mAs at 71 ∼ 80 kV
0.5 ∼ 80 mAs at 81 ∼ 90 kV
0.4 ∼ 60 mAs at 91 ∼ 100 kV
0.4 ∼ 60 mAs at 101 ∼ 120 kV
◇ Cool-down period : 2 min. at maximum output
◇ Maximum deviation ; - kV : ± 7%
- mAs : ± 10%
◇ Filtration ;
- X-ray tube : min. 0.8 mmAleq.
- X-ray tube assembly : min. 1.5 mmAleq.
- Beam limiting device : min. 1.3 mmAleq.
- X-ray source assembly : min. 2.8 mmAleq.
-------------------------------------------------------------
AJEX240H / Operating instructions -33-
AXR-240H-MN01 (E) / Rev. 03