2) ERROR code E02 ; kV or mA overload error
a) Press enter key, then error mode will be returned normal radiographic
mode. If error mode is not returned normal radiographic mode,
contact AJEX immediately.
b) Releasing an exposure.
If error code E02 is displayed again on the mAs display, contact
AJEX immediately.
☞ Refer to page 28. 6-3. Service
3) ERROR code E03 ; kV or mA feedback undetected error
a) Press enter key, then error mode will be returned normal radiographic
mode. If error mode is not returned normal radiographic mode,
contact AJEX immediately.
b) Releasing an exposure.
If error code E03 is displayed again on the mAs display, contact
AJEX immediately.
☞ Refer to page 28. 6-3. Service
4) ERROR code E04 ; Unintended X-ray exposure finish error
CAUTION
The exposure ends immediately after the operator stops pressing the
exposure release switch. ( deadman's function )
Therefore, press the exposure release switch to exposure condition
(i.e. second position) and hold until the exposure is finished.
-------------------------------------------------------------------------
AJEX240H / Operating instructions -31-
AXR-240H-MN01 (E) / Rev. 03