EasyManua.ls Logo

Alcedo B21 - Immunity Test Levels by Frequency (Cont.)

Alcedo B21
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
Guidance and manufacturer’s declaration - electromagnetic immunity
Pulse Modulation
217 Hz
b)
1720
1845
1970
2 0.3 281700 – 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3
4, 25;UMTS
Pulse Modulation
217 Hz
b)
2450 2 0.3 282400 – 2570
Bluetooth,
WLAN
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse Modulation
217 Hz
b)
5240
5500
5785
0.2 0.3 95100 – 5800
WLAN 802.11
a/n
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC
61000-4-3.
a ) For some services, only the uplink frequencies are included.
b ) The carrier shall be modulated using a 50% duty cycle square wave signal.
c ) As an alternative to FM modulation. 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.