Guidance and manufacturer’s declaration - electromagnetic immunity
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT,
and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance.
Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following
equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST
LEVEL in V/m.