Electromagnetic immunity
40
Immuni-
ty tests
IEC
60601
test
level
Com-
pliance
level
Electromagnet-
ic environment
– guidelines
Electro-
static dis-
charge
(ESD) in
accor-
dance
with IEC
61000-4-2
± 6 kV
contact
dis-
charge
± 8 kV
air dis-
charge
Not ap-
plicable
± 8 kV
air dis-
charge
Floors should
be made of
wood or con-
crete or be
covered with
ceramic tiles.
If the floor is
covered with
synthetic mate-
rial, the relative
humidity must
amount to at
least 30 %.
Fast
transient
electrical
bursts
in accor-
dance
with IEC
61000-
4-4
± 2 kV
for
power
lines
±1 kV
for in-
put and
output
lines
Not ap-
plicable
The quality of
the supply volt-
age should cor-
respond with
a typical store
or hospital
environment.
Surges
in accor-
dance
with IEC
61000-
4-5
± 1 kV
sym-
metrical
voltage
± 2 kV
asym-
metrical
voltage
Not ap-
plicable
The quality of
the supply volt-
age should cor-
respond with
a typical store
or hospital
environment.