Electromagnetic immunity
43
Immu-
nity
tests
IEC
60601
test
level
Com-
pliance
level
Electromagnetic
environment
– guidelines
Con-
ducted
HF
distur-
bances
in accor-
dance
with IEC
61000-
4-6
3 Vrms
150
kHz to
80
MHz
3 Vrms d = 1.2 √P
Emit-
ted HF
distur-
bances
in accor-
dance
with IEC
61000-
4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m d = 1.2 √P
80 MHz to
800 MHz
d = 2.3√P
800 MHz to
2.5 GHz
where P is the
nominal output
of the transmit-
ter in watts (W)
according to the
specifications of
the transmitter
manufacturer
and d is the
recommended
safety clearance
in meters (m).
According to an
on-site survey
a
,