UV Intensity Sensor Calibration Resetting
In case of a new installation or when the lamp is replaced with a new one, it is important to
reset the sensor calibration settings. This must also be performed in case a new sensor is being
fitted to the system.
Following is the procedure:
1. Ensure that the system and lamp is powered up and there is liquid flowing through the
reactor.
2. Let the system run for at least 1 hour till it reached a steady state.
3. To reset the UV Intensity Sensor to 100%, press the Function key on the controller for 7
seconds.
4. Once confirmed, the controller will beep once and the screen will display 100%.