AVF7000 Operation Manual OP7000.1.1.7.R12ENG
23
11.5 Amplification Gain Settings
11.6 Temperature / Pressure Calibration
Password
Set by and keys. Enter password XXXXX then press long for 3s to enter temperature/
pressure calibration settings (If necessary, please contact ALIA).
Code
00000
Temperature Input Mode
There are 2 options: Manual and Auto. Manual mode, achieving temperature/pressure
compensation with setting value in window Temperature. Auto mode, PT1000 should be connected
externally. Press
to enter next setting interface.
Temp.Measure
Auto
Pressure Input Mode
There are 2 options: Manual and Auto. Manual mode, achieving temperature/pressure
compensation with setting value in window Pressure. Auto mode, a pressure sensor should be
connected externally. Press
to enter next setting interface.
Pressure Measure
Auto
Temp Low Point Calibration(Not proceed unless necessary, otherwise low accuracy)
Connect a standard 1000Ohm resistance externally with the circuit board then press for 3s and
release until value 1000.00 flashes; after this press
for 3s again until value 1000 Ohm stops
flashing then release. Temp Low point calibration is finished. Press
to enter next setting
interface.
Temp. Low Trim
1000.00 ohm
Temp High Point Calibration(Not proceed unless necessary, otherwise low accuracy)
Connect a standard 2500Ohm resistance externally with the circuit board then press for 3s and
release until value 2500.00 flashes; after this press
for 3s again until value 2500 Ohm stops
flashing then release. Temp High point calibration is finished. Press to enter next setting
interface.
Temp. High Trim
2500.00 ohm
Pressure Low Point Calibration(Not proceed unless necessary, otherwise low accuracy)
Connect a pressure sensor externally and input 0kpa then press for 3s and release until value
0.00000 is flashing; after this press
for 3s again and release until value 0.00000 stops flashing.
Pressure Low point calibration is finished. Press
to enter next setting interface.
Pre. Low Trim
0.000 kpa
Pressure High Point Calibration(Not proceed unless necessary, otherwise poor accuracy)
Connect a pressure sensor externally and input pressure, such as 1000kpa, and press
for 3s
and release until value 1000.00 is flashing; after this press
for 3s again and release until value
1000.00 stops flashing. Pressure High point calibration is finished. Press
to enter next setting
interface.
Pre. High Trim
1000.00 kpa
+
Password
Press and to enter password 00062, then hold key for 3s to enter channel setting
interface.
AMP.Channel
CH3
Amplification Gain Settings
Press and to set amplification gain: CH1, CH2, CH3; CH_1 means amplifying 10
1
times on
amplification factor, it’s usually used to measure liquid. CH3 means amplifying 10
3
times on
amplification factor, it’s usually used to measure gas. CH2 means amplifying 10
2
times on
amplification factor, it’s usually used to make adjustment according to fluid condition and
interference strength. Press
SETSET
twice to return to normal interface.
Code
00000
+