AUTOCLAVE
(Steam Sterilizer)
CL- 32S / 32L
40S / 40M / 40L
Instruction Manual
Make sure to read this manual to use this product safely.
Place this manual in a convenient location for easy reference.

Questions and Answers

  • D
    Daniel FosterSep 10, 2025
    Why does evacuating residual air not stop at 100C in my ALP CL- 32L Laboratory Equipment?
    • T
      Thomas VanceSep 10, 2025
      If the ALP Laboratory Equipment does not stop evacuating residual air around 100? and the temperature doesn't rise, it could be due to a breakage of the exhaust valve or because the autoclave is installed over 1000m above sea level. Contact your local distributor, because this autoclave cannot be used over 1000m above sea level.
  • H
    Heidi MillerAug 23, 2025
    What to do if the lid lock lever on my ALP CL- 32L cannot be released?
    • P
      perryelizabethAug 23, 2025
      If the lid lock lever of the ALP Laboratory Equipment cannot be released, ensure the power switch (breaker) is turned on. If it is on, and the TEMPERATURE DISPLAY indicates above 80° after operation complete or the COMPLETE LAMP lights up after operation, press the STOP KEY to release the lid lock.
  • D
    Doris KingAug 25, 2025
    Why does the breaker trip immediately when I turn on my ALP Laboratory Equipment?
    • D
      davidsonalejandroAug 25, 2025
      If the breaker trips immediately after powering on the ALP Laboratory Equipment, it could be due to electrical leakage (possibly from insulation deterioration or water leakage at the heating element) or a short circuit (in the heating element or elsewhere). In any of these cases, contact your local distributor.
  • R
    Robert DavisAug 27, 2025
    Why is the safety valve activating on my ALP CL- 32L?
    • G
      Grant ClementsAug 28, 2025
      If the safety valve activates on the ALP Laboratory Equipment (steam blowing from the bottom plate), it may be because residual air is not evacuated enough, causing abnormal pressure, or the chamber pressure surpassed set pressure of safety valve due to the malfunction of temperature sensor. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. If the issue persists, contact your local distributor.
  • B
    Brett ParksAug 29, 2025
    Why isn't sterilization completing in my ALP CL- 32L Laboratory Equipment even at the set temperature?
    • M
      Michael Hardy DVMAug 29, 2025
      If the ALP Laboratory Equipment reaches the set temperature but sterilization isn't completed because the chamber doesn't fill with saturated steam, it could be due to several factors: * Too much capacity of objects in chamber. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. * Shortage of sterilization time. Increase sterilization time appropriately, especially for liquid objects or large quantities. * Residual air inside container prevents penetration of steam. If using disposable plastic bags, ensure they are filled to less than 60% capacity and have an open mouth. * Air evacuation is insufficient due to bent exhaust hose. Make sure the exhaust hose is not bent.
  • C
    Corey ThompsonNov 25, 2025
    What does Er1 mean on my ALP Laboratory Equipment and how do I fix it?
    • C
      Christopher CastroNov 25, 2025
      If 'Er1' appears on the ALP Laboratory Equipment display with an alarm and the operation stops, it could be due to several reasons: the autoclave is working without water in the chamber, the drain valve is open, there is breakage of water lack preventive device or over pressure detector, over pressure detector activated, residual air in the capacious container or disposable plastic bag swelled or air evacuation is insufficient due to bent exhaust hose. To address this, turn off the power switch and let the autoclave sit for at least one hour to cool down, then pour water into the chamber and close the drain valve. In case of objects being put in a disposable plastic bag, quantity of objects to be put inside disposable plastic bag shall be less than 60% of capacity of the chamber and disp...
  • K
    Keith RomeroSep 18, 2025
    Why does drainage not stop from my ALP CL- 32L Laboratory Equipment?
    • B
      Brittany EdwardsSep 18, 2025
      If drainage from the ALP Laboratory Equipment doesn't stop, it suggests a breakage of the drain valve, possibly due to glass or other objects damaging the valve's Teflon ball. Contact your local distributor.
  • C
    Carrie FisherSep 22, 2025
    What should I do if the chamber temperature of my ALP CL- 32L Laboratory Equipment doesn't rise?
    • S
      smedinaSep 22, 2025
      If the chamber temperature of the ALP Laboratory Equipment does not rise after pressing the START KEY, it could be due to breakage of the heating element or burning of the wire connected to the heating element. Contact your local distributor.
  • L
    Lisa ThompsonSep 8, 2025
    What does error code Er3 mean on my ALP CL- 32L Laboratory Equipment?
    • R
      Richard WalkerSep 8, 2025
      If 'Er3' is shown on the ALP Laboratory Equipment display with an alarm, it indicates that the chamber temperature rose 2? higher than the setting temperature during sterilization. Ensure the distance between the chamber wall and objects is more than 10mm. Oleaginous objects must not be sterilized and if the problem persists, contact your local distributor.
  • E
    Erica HarmonSep 5, 2025
    Why is there steam leaking from the lid of my ALP CL- 32L?
    • K
      Kristin WatsonSep 5, 2025
      If there's steam leaking from the lid of the ALP Laboratory Equipment, it could be due to deterioration of the lid gasket (cracks or hardening) or misalignment of the lid and chamber. Contact your local distributor to resolve this issue.

Summary

For the safety use

Preamble and Safety Symbols

Introduces safety importance and defines warning symbols.

Autoclave Usage Safety

Covers safe operation and potential hazards.

Installation Safety

Outlines safety measures for installing the autoclave.

Cleaning Safety Precautions

Details safety measures for cleaning the autoclave.

Operational Safety Notes

Provides further safety aspects and warnings during autoclave use.

Power Failure Safety Procedures

Details actions and precautions during power interruptions.

Part names

Installation

Power Cable and Grounding

Electrical connection and protective grounding instructions.

Operation Method

Basic Operation Flowchart

Step-by-step guide for standard operation.

Starting the Operation

Procedure to initiate the sterilization cycle.

Operation Completion and Cooling

Steps after cycle completion, including cooling.

Safety Device

Safety Valve Functionality

Description of the safety valve and its activation mechanism.

Over Pressure Detection System

How the system detects and responds to excess pressure.

Water Lack Prevention Device

Mechanism to prevent operation without sufficient water.

Trouble shooting

Error Code Reference

A list of error codes and their meanings.

Troubleshooting Common Issues

Detailed guide for diagnosing and resolving operational problems.

Maintenance

Specifications

ALP CL- 32L Specifications